- Heated Incubator
- 10-Step Programmable
- Temperature Range ambient +5 °C to +70 °C
- Chamber Volume:
-
- 50 liters (JSGI-050T)
- 100 liters (JSGI-100T)
- 150 liters (JSGI-100T)
- 250 liters (JSGI-100T)
-
General Incubator, Programmable
- Heated Incubator
- 10-Step Programmable
- Temperature Range ambient +5 °C to +70 °C
- Chamber Volume:
-
- 50 liters (JSGI-050T)
- 100 liters (JSGI-100T)
- 150 liters (JSGI-100T)
- 250 liters (JSGI-100T)
-
Categories General Incubator, General Lab Equipment, Incubator
Brand: JSR
Specifications
| Heating System | Forced Air Convection |
| Temp. Range | ambient+5 oC ~ 70 oC |
| Accuracy | ± 0.2 oC at 37.0 oC (JSGI-050T) |
| Uniformity | ± 0.5 oC at 37.0 oC (JSGI-050T) |
| Controller | JIC-CON 5th Generation Programmable Temperature Controller |
| Program | 1 Pattern, 10 Steps, 1 ~ 999 (or unlimited) times repreat |
| 3-Zone Temperature Control Algorithm | |
| 3-Point Temperature Calibration | |
| 1-Point Temperature Offset Calibration | |
| Sensor | Class A Pt-100 Ω Sensor |
| Safety | Over-Temperature Cut-Off |
| Over Current Cut-Off | |
| Material | Body – Epoxy Powder Coated Steel |
| Chamber – Stainless Steel 304 | |
| Inner Door – 5 mm thick Tempered Safety Glass | |
| including PVC coated wire shelf | |
| optional perforated stainless steel shelf | |
| Electrical Data | 220±10% VAC 50/60Hz 1-Phase |
| Plug config. | CEE 7/7 Schuko or BS 1363 |
Features
- One pattern 10 steps program – User can compose 10 steps of varying temperature program and can save it in a memory. Composed program can be repeated 1 ~ 999 times or unlimited number of time.
- 3 group zone PID algorithm – Complete temperature range is divided into three zones (LOW/MIDDLE/HIGH) and each ZONE has its individual PID Algorithm to provides precise and superior temperature control characteristic of whole temperature range.
- 3 Point calibrations – Complete temperature range is divided into three zones (LOW/MIDDLE/HIGH) and each ZONE has its individual PID Algorithm to provides precise and superior temperature control characteristic of whole temperature range.
- 1 Point off set calibration – Provides simple and easy calibration tool to calibrate chamber at ONE SPECIFIC TEMPERTURE POINT of frequent use.